The number of simultaneously verified devices: from 1 to 4 with the same ranges
Pressure setting: pneumatic pressure up to 21 MPa and hydraulic pressure up to 70 MPa, vacuum (residual pressure) 1 kPa.
Pressure standards:
−Metran series electronic pressure calibrators;
− Metran-500 Air series pneumatic pressure calibrators;
− deadweight pressure gauges, barometers
Standards for output signal measurement:
− Metran series calibrators and multimeters;
− voltmeter with a measure of electrical resistance
Pressure/vacuum setting methods:
−manual, automated, automatic
Power supply: pressure transmitters with a voltage of 24 and 36 V, electric contact pressure gauges – 24 V
Test bench power supply:
− electric power supply 220 V, 50 Hz;
− pneumatic power supply from an external source (pneumatic network, compressor, power supply system, vacuum pump)
Table dimensions: no more than 1600x1450x810 mm
Metrology bench is used as a ready-made separate workplace or as part of metrological laboratories, and is used for verification or calibration of pressure transmitters, pressure gauges (including signaling pressure gauges), pressure switches, hydrostatic level transmitters, pressure transmitters as part of variable differential pressure flowmeters.
The choice of the bench configuration is made in accordance with individual requirements for the degree of automation, taking into account the features of the park of verified transmitters.
Standard solutions for pressure instrument verification benches are also available on request including a mandatory short-list of equipment for verification and calibration of pressure instruments.
The possibility of simultaneous verification of up to 4 pressure transmitters or pressure gauges